dc.contributor.author | Albrecht, Roland | |
dc.contributor.author | Kunz, Andres | |
dc.contributor.author | Voelckel, Wolfgang G | |
dc.date.accessioned | 2016-04-07T06:54:15Z | |
dc.date.available | 2016-04-07T06:54:15Z | |
dc.date.issued | 2015-02 | |
dc.identifier.citation | Albrecht, R. et al., (2015) Airplane transport isolators may loose leak tightness after rapid cabin decompression. Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine, 23:16 | nb_NO |
dc.identifier.uri | http://hdl.handle.net/11250/2384356 | |
dc.description | This article was originally published in Scandinavian Journal of Trauma, Resuscitation
and Emergency Medicine, and distributed under the terms of the Creative
Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and
reproduction in any medium, provided the original work is properly credited. | nb_NO |
dc.description.abstract | Air medical transport of patients suffering of highly infectious diseases is typically performed employing portable isolation chambers. Although the likelihood of decompression flight emergencies is low, sustainability of the devices used is crucial. When a standard isolation unit was subjected to an explosive cabin decompression of 493 hPa, simulating a 32808 ft flight level accident, leak tightness of the unit was lost due to rupture of the bag caused by over expansion. When the pressure chamber experiment was repeated with a modified unit, distension was minimized by an additional compensation air bag, thus ensuring leak tightness. | nb_NO |
dc.language.iso | eng | nb_NO |
dc.publisher | BioMed Central | nb_NO |
dc.rights | Navngivelse 3.0 Norge | * |
dc.rights.uri | http://creativecommons.org/licenses/by/3.0/no/ | * |
dc.subject | luftambulanse | nb_NO |
dc.subject | air medical transport | nb_NO |
dc.subject | infectious diseases | nb_NO |
dc.subject | isolation chamber | nb_NO |
dc.subject | smittevern | nb_NO |
dc.title | Airplane transport isolators may loose leak tightness after rapid cabin decompression | nb_NO |
dc.type | Journal article | nb_NO |
dc.type | Peer reviewed | nb_NO |
dc.rights.holder | © 2015 Albrecht et al. | nb_NO |
dc.subject.nsi | VDP::Medical disciplines: 700 | nb_NO |
dc.source.volume | 23 | nb_NO |
dc.source.journal | Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | nb_NO |
dc.identifier.doi | 10.1186/s13049-015-0090-6 | |